beautypg.com

INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 238

background image

6 - 4

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

Figure 6-2 Data File Information Page

During the Calibration run, enter the concentration level of the calibration standard
and select the corresponding units.

A description can be entered to help identify the sample.

Open the Calibrate Function using the Toolbar Icon or the drop down menu. Select
Survey as the file type and select OK. The File selection dialog box will be
displayed. Locate and select the method that was used to collect the calibration
runs.

Use the Browse button in the Data Files section to select and open the data file
collected during the calibration run. The file(s) will be used to build the target library
template. See

Figure 6-3

.

Highlighting a data file and pressing the Display button will display the calibration
run.

NOTE: The best way to work with the graphical data is to expand the plot to full

screen.

Figure 6-3 Calibration Window