HAPSITE Smart Plus Operating Manual

The quantitation sequence method must be defined. This includes the
definition of:

the Data file (where the data will be stored)

the Selected Ion Method parameters (which defines the masses to be
monitored and the elution times expected, as well as other parameters of
the mass spectrometer operation)

the Library Search parameters (which designates which library file to
access, individual and collective peak search, integration parameters, and
various reporting options)

the Inlet method file (which defines the temperatures, timing, and other gas
chromatograph and head space parameters)

the Tune/calibration file (which defines the mass spectral response to the
target compounds and the internal standards)

The HAPSITE must be tuned, either with the AutoTune or Manual Tune
program. (Refer to

Chapter 7, Tune

.)

The HAPSITE must be calibrated to the target compounds. (See

Chapter 12,

Target Compound Methods

.)

WARNING

Temperature settings are critical to get the proper results.
Only trained users should modify these.