HAPSITE Smart Plus Operating Manual

Figure 6-5 Grab Spectrum for Template Window

The compound is displayed in the Analytes in Library list and the extracted mass
values are displayed. A compound name can be entered and the extracted mass
list can be "cleaned up" for low intensity masses and more than one VOC in the
Survey run. To edit the extracted mass list, all masses with intensity below 15%
should be deleted. To delete a mass, click once on the mass in the Extracted Mass
Peaks
list and press the delete key. After adding all analytes to the template, the
library can now be calibrated by checking the Selection box for the data file,
selecting Calibrate Library, and then pressing Start.

The library should be saved at this time, the Save Library button under Method will
display the dialog box to save the library. The library is linked to the method and
can be saved using any name.

Additional calibration points can be added to the curve by using the Browse button
to select additional data files. The selected files can then be processed as
described above, and saved to the library. See

Figure 6-6

.