HAPSITE Smart Plus Operating Manual

is applied to the top elbow, the elbow should not slip on the Sample Loop. If the
elbow moves, the Loop is not properly seated. Reseat Loop and tighten slightly
more. See

Figure 2-54

.

Figure 2-54 Wrench tighten

8

Snap on the black cover labeled SAMPLE LOOP. The cover will fit snugly over
the Sample Loop. See

Figure 2-55

.

NOTE: If the cover does not easily fit over the Sample Loop, do not force it. Check

to ensure the Loop is correctly installed with the Loop fully seated into both
elbows and the nuts properly tightened for a secure fitting.

Figure 2-55 Sample Loop Cover Installed