HAPSITE Smart Plus Operating Manual

Procedure

1

If an Analyze method is going to be run after Survey, verify that the appropriate
sample configuration (i.e., concentrator) is installed.

2

When powered on or taken out of Extended Standby, the HAPSITE will
automatically start preparing an Survey and Analyze method if the probe is
attached. Refer to

Section 3.1, Starting the HAPSITE in Portable Mode, on

page 3-1

.

3

Verify that the desired Analyze method is listed under the Survey method. In

Figure 3-30

, the Analyze method is Air_Tri-Bed_PPM_Standard.

Figure 3-30 Analyze Method

4

A SYSTEM READY message will appear with instructions to press Survey or
Analyze when the HAPSITE Smart Plus is ready to sample. See

Figure 3-31

.

Figure 3-31 System Ready