beautypg.com

INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 209

background image

5 - 3

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

4

As part of the HAPSITE’s preparation, it will run AutoTune. If AutoTune
successfully runs and completes, the AutoTune OK message will appear. Click
OK. See

Figure 5-3

. If AutoTune fails, see

Section 7.4, AutoTune Failure, on

page 7-6

. When the final results message is displayed, click OK.

Figure 5-3 AutoTune Complete

5

The HAPSITE will check pressures and automatically heat all necessary zones
to the setpoint temperatures specific to the selected method. Progress will be
indicated by a bar graph.

6

Click the RUN METHOD icon. Click on the Survey folder. Click on the folder of
the desired configuration, and then click on the method folder. See Steps 2-4 of

section 5.3

.

CAUTION

Do not touch the sample with the Probe. Do not allow
liquids to enter the Probe.

7

Click the RUN button on the pop-up window or from the control panel on the
screen. See

Figure 5-4

.

Figure 5-4 Run Button

8

Sample the background for one minute and note the TIC.

9

Hold the Probe over the sample of interest for up to 1 minute. If the TIC begins
to rise, slowly back the Probe away from the sample. If the TIC does not rise
after a full minute, move the probe away from the sample.

9a

A peak may appear if the compound present is >10 ppm. A compound
identification may also be present on the HAPSITE screen.

NOTE: No response may indicate either the compound present is less than the

detection limit, or that no detectable compound is present.