12.3.3 Peak Search
Search . . . . . . . . . . . . . . . . . . . . . . . Performs a peak detection and integration on
the selected files, producing the quantitation
report.
Recalculate . . . . . . . . . . . . . . . . . . . Integrates analyte peak areas and
recalculates the concentrations without
performing a peak search. Typically used
following manual editing of peak baseline
points.
Start . . . . . . . . . . . . . . . . . . . . . . . . . Initiates the Search or Recalculation. See
Figure 12-5
.
Figure 12-5 Analytes Display
12.3.4 Analytes
Analytes in Library . . . . . . . . . . . . . Displays the analytes in the library.
Analytes in File . . . . . . . . . . . . . . . . Displays the analytes in the currently
displayed or selected file.
12.3.5 Reports
Calibration Response Table . . . . . Report that displays the response factor and
curve statistics based on the selected curve
type.
Calibration Report . . . . . . . . . . . . . Report that displays the area fit and purity for
the calibration standards.
View Report. . . . . . . . . . . . . . . . . . . Displays the selected calibration report.
# . . . . . . . . . . . . . . . . . . . . . . . . . . . . Shows the analyte number in the library.
D. . . . . . . . . . . . . . . . . . . . . . . . . . . . Data Reference: lists the reference to the
data file in which the analyte was found.