16 - 32

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

7

Save the rubber “O” ring assembly and container to dispose of the old NEG
Pump.

8

Place the new aluminum gasket on the new NEG Pump. See

Figure 16-31

.

NOTE: Inspect the manifold face, ionizer and manifold inside for nicks,

residue, and scratches.

NOTE: Write down the Serial Number of the new NEG Pump.

Figure 16-31 Aluminum Gasket on NEG Pump

9

Insert the new NEG Pump and aluminum gasket into the manifold. See

Figure 16-32

.

Figure 16-32 Installing New NEG Pump