2 using the front panel display to detach – INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 493

15 - 19
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
6
When the Detach procedure is complete, a message will appear confirming a
successful detachment. See
.
Figure 15-27 Detach Successful
NOTE: To place the Service Module in storage see
.
15.8.2 Using the Front Panel Display to Detach
1
Touch Menu or push MENU.
2
Touch the PROBES & SM icon. Alternately, push the STAT button until the
PROBES & SM page is displayed.
3
Touch the DETACH SM button. Alternately, using the arrow keys, highlight the
DETACH SM button. Push OK SEL
4
A bar graph displaying the progress of the Detach procedure will appear on the
front panel of the HAPSITE.
5
The message Service Module Found, Not Attached will be displayed on the
screen.
CAUTION
Wait until the Turbo Speed reads zero before physically
removing the HAPSITE from the Service Module.
6
To place the Service Module in storage, see
section 15.10, Storing the Service
15.9 Physically Removing the HAPSITE from the Service Module
The HAPSITE cannot be physically removed from the SM until a "Detach SM" has
been performed. Refer to
section 15.8, Detaching the HAPSITE, on page 15-17
before physically removing the HAPSITE from the Service Module.
After the HAPSITE has been detached from the Service Module, release the
latches on each side of the Service Module.
HINT: To continue using the HAPSITE, place a charged battery in the HAPSITE
or connect exterior power before removing the HAPSITE from the docked
position on the Service Module.