5.2.1 Quick Reference SOP — Running Survey Mode

1

Either power on the HAPSITE or take the HAPSITE out of Extended Standby.

2

Double-click the Plus IQ software icon.

3

Wait for heaters to reach the set temperatures.

4

Select the desired method. See Steps 2-4 of

section 5.3 on page 5-7

.

5

Click the RUN button in the pop-up window.

6

Sample background for one minute.

CAUTION

Do not touch the sample with the Probe. Do not allow
liquids to enter the Probe.

7

Hold the Probe over the sample for one minute or until observing a response.

8

Press Stop to stop the sampling and save the data.

NOTE: This is a tentative identification. For more positive identification, run an

Analyze (GC/MS) method.