4 operating specifications – INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 461
14 - 3
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
14.4 Operating Specifications
Environmental Temperature . . . . . . . Operating: 5 to 45°C, 0-95% RH
(non-condensing)
Storage: -40 to 70°C (stored dry)
Water Sample
Temperature . . . . . . . . . . . . . . . . 5 to 30°C
Above 26°C, Dry Purge may be extended
Conductivity . . . . . . . . . . . . . . . . ionic conductivity is used to sense the proper
water level in the probe head before starting
the analysis. Therefore, the water sample
must exhibit conductivity before the analysis
can begin. Very pure water, such as
deionized or distilled water, may require the
addition of a salt before the analysis
sequence will start.
pH Range . . . . . . . . . . . . . . . . . . 2-11
Probe Depth . . . . . . . . . . . . . . . . . . . minimum, 5.5 in (140 mm)
maximum, 18 in (457 mm)
Heated Zones
Probe . . . . . . . . . . . . . . . . . . . . . 30 to 60°C
Oven . . . . . . . . . . . . . . . . . . . . . . 30 to 60°C
Transfer Line. . . . . . . . . . . . . . . . 30 to 60°C
80°C during Flush Mode
(External Power Only)
Practical Quantitation Limit. . . . . . . . ppt to ppb, based on HAPSITE configuration
and method parameters
Internal N2 Carrier Gas: . . . . . . . . . . Standard HAPSITE (770 kPa) N2
disposable can
Power Requirements
DC Power . . . . . . . . . . . . . . . . . . 24 V(dc) ±5% via HAPSITE Power Adapter
80 W maximum
Battery Power . . . . . . . . . . . . . . . Standard 19.2 V(dc) HAPSITE battery. Flush
mode not recommended on battery power.
Software . . . . . . . . . . . . . . . . . . . . . . HAPSITE Plus IQ Software