HAPSITE Smart Plus Operating Manual

7.6.1 Adjusting Base Peak Gain

1

Double-click on the Plus IQ icon. See

Figure 7-27

.

Figure 7-27 Plus IQ Icon

2

Double-click on the Tune icon. See

Figure 7-28

.

NOTE: Manual Tune can only be opened in Advanced User Mode.

Figure 7-28 Manual Tune Icon

3

Select default.tun file and press OK. See

Figure 7-29

.

Figure 7-29 Select default.tun File

4

Wait for the automated process of opening the tune to complete.

5

Check the Base Peak Gain (BPG). Is the value between 0.8 and 2.0? See

Figure 7-30

and

Figure 7-31

.