4 analyze (gc/ms) mode with loop – INFICON HAPSITE Smart Plus Chemical Identification System User Manual

Page 216

background image

5 - 10

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

5.4 Analyze (GC/MS) Mode with Loop

This method is used for analyzing samples with concentration levels in the low part
per million (ppm) to high part per billion (ppb) range. This can be used after Survey
to detect and/ or confirm unknowns in the environment. See

Chapter 6, Methods

for additional information on (GC/MS) methods.

WARNING

Remember that the Survey method must be followed by
an Analyze (GC/MS) method in order to give a positive
identification of the unknown.

This section will describe the process required to run a sample with the Sample
Loop installed. Make sure that the Sample Loop is installed before turning on or
taking the HAPSITE out of Extended Standby. Refer to

Section 2.4, Basic

Assembly, on page 2-5

and

Section 2.10.1, Installing the Sample Loop, on page

2-31

.

1

Double-click the Plus IQ icon to open the software. See

Figure 5-13

.

Figure 5-13 Plus IQ Icon

2

When powered on or taken out of Extended Standby (refer to

section 3.1 on

page 3-1

or

section 3.8.1 on page 3-63

), the HAPSITE will automatically start

preparing the start up Loop method. If this method is not the desired method,
see

Section 5.3, Selecting a New Method, on page 5-7

.

3

As part of the preparation progress, it will heat all necessary components,
check pressures and run an AutoTune if necessary. Progress of the preparation
will be shown as a bar graph on the Laptop screen. If AutoTune fails, see

Section 7.4, AutoTune Failure, on page 7-6

. When the final results message is

displayed, see

Figure 5-14

, click OK.

Figure 5-14 AutoTune Complete