5.4 Analyze (GC/MS) Mode with Loop
This method is used for analyzing samples with concentration levels in the low part
per million (ppm) to high part per billion (ppb) range. This can be used after Survey
to detect and/ or confirm unknowns in the environment. See
for additional information on (GC/MS) methods.
WARNING
Remember that the Survey method must be followed by
an Analyze (GC/MS) method in order to give a positive
identification of the unknown.
This section will describe the process required to run a sample with the Sample
Loop installed. Make sure that the Sample Loop is installed before turning on or
taking the HAPSITE out of Extended Standby. Refer to
Section 2.10.1, Installing the Sample Loop, on page
1
Double-click the Plus IQ icon to open the software. See
.
Figure 5-13 Plus IQ Icon
2
When powered on or taken out of Extended Standby (refer to
or
), the HAPSITE will automatically start
preparing the start up Loop method. If this method is not the desired method,
see
Section 5.3, Selecting a New Method, on page 5-7
3
As part of the preparation progress, it will heat all necessary components,
check pressures and run an AutoTune if necessary. Progress of the preparation
will be shown as a bar graph on the Laptop screen. If AutoTune fails, see
Section 7.4, AutoTune Failure, on page 7-6
. When the final results message is
displayed, see
, click OK.
Figure 5-14 AutoTune Complete