HAPSITE Smart Plus Operating Manual

7.4 AutoTune Failure

Occasionally, AutoTune may fail. Follow the procedures outlined below before
continuing.

1

When AutoTune fails, check IS canister pressure and expiration date. Replace
if necessary.

2

Check reservoir pressure to ensure the reservoir pressure is at least 400 kPa.
Replenish if necessary by inserting a full can of Nitrogen (>400 kPa).

3

Rerun AutoTune by rebooting the HAPSITE.

4

If AutoTune fails again, select File >> View Tune Reports on the Laptop. See

Figure 7-9

.

Figure 7-9 Selecting View Tune Reports