NOTE: In addition to the standard 15 minute methods, there are 25 minute
methods for the Sample Loop and Concentrators. These methods end in
Extended in place of the Standard portion of the name. The analysis of
Toxic Industrial Chemicals is the suggested use for these methods. The
extended analysis time provides improved resolution of multiple
compounds and elution of higher boiling point compounds.
8.5.2 Loading a Default Method to the HAPSITE
1
From the Default Methods Menu, select the default method to load. Refer to
Figure 8-14, HAPSITE Default Methods, on page 8-8
.
HINT: Refer to
Section 8.5.1, Locating Default Methods, on page 8-8
the default methods.
2
Press the Load button. See
Figure 8-14, HAPSITE Default Methods, on page
.
3
Press the Save button at the bottom of the Method Editor Description window.
See
.
NOTE: The method file name is the default method name with a two digit
number appended (e.g., 01). If the two digits are not desired, then
remove them from the file name before pressing the Save button.
Figure 8-15 Default Methods Method Editor Description Window