HAPSITE Smart Plus Operating Manual

Max OK Target Percentage . . . . . . Displays the maximum percentage required

for the mass peak to meet the OK High
criteria.

Base Peak . . . . . . . . . . . . . . . . . . . . Displays the base peak that the mass peak

percentage is measured against.

7.5.7 Mass Calibration Status

The dark gray Mass Calibration Status panel displays the status of the last Mass
Calibration. See

Figure 7-23

. The Mass Calibration Status panel can be displayed

by Right Mouse Button (RMB) click in the Peak Scan window, or by using the Tune
drop down menu and selecting the Mass Calibration Status.

Figure 7-23 Mass Calibration Status

Columns can be displayed or hidden with a Right Mouse Button click on the column
headings. See

Figure 7-24

.

Figure 7-24 Mass Calibration Menu

Mass . . . . . . . . . . . . . . . . . . . . . . . . Mass number.

Low%. . . . . . . . . . . . . . . . . . . . . . . . Minimum percentage for peak status OK.

High% . . . . . . . . . . . . . . . . . . . . . . . Maximum percentage for peak status OK.

Base. . . . . . . . . . . . . . . . . . . . . . . . . Reference mass for peak percentage

calculations.