7 - 3

IP

N 07

4-

47

2-

P1

C

HAPSITE Smart Plus Operating Manual

7.2.1 Starting AutoTune from the Manual Tune Screen on the Laptop

Computer

1

Double-click on the Plus IQ icon. See

Figure 7-2

.

Figure 7-2 Plus IQ Icon

2

Double-click on the Tune Icon. See

Figure 7-3

.

HINT: Only the Advanced Access level can run Manual Tune.

Figure 7-3 Manual Tune Icon

3

Wait until the EM and Emission buttons in the control panel turn green, then
click the Tune icon. See

Figure 7-4

.

Figure 7-4 Tune Icon

CAUTION

Adjusting other parameters without proper training may
damage the instrument.

4

Allow the AutoTune to run to completion and click OK. Close the Manual Tune
after receiving the message Final Results are on the LCD. See

Figure 7-5

.

Figure 7-5 AutoTune Complete