2.11.3 VX / R-33 Conversion Tube
This procedure describes the steps required to prepare the HAPSITE to sample for
VX or R-33 using the Conversion Tube. To detect VX or R-33, you must insert the
Conversion tube following the guidelines below.
The process of detecting VX or R-33 on the HAPSITE requires the conversion of
VX or R-33 (high boiling point chemicals) to the G analog. The VX or R-33 molecule
is broken at the sulfur bond when it comes in contact with a silver fluoride pad. The
result is the formation of a volatile chemical ethyl methyl phospho no fluoridate in
the case of VX, or isobutyl methyl phospho no fluoridate in the case of R-33. These
compounds are detected by the HAPSITE as VX-G or R-33-G.
NOTE: Theoretically, other G agents can be detected with the VX conversion pad
in place. However, if other G agents are suspected, the sample should also
be run without the conversion tube in place.
WARNING
Sulfur mustard cannot be detected with the conversion
tube in place.
2.11.3.1 VX / R-33 Conversion Tube Installation
1
section 2.11.1, Probe Nut Assembly, on page 2-42
to ensure proper
ferrule orientation in the probe nut.
WARNING
Correct ferrule placement is critical to ensure a leak-free
fit around the conversion tube.
2
Loosen the nut on the end of the probe approximately 1/4 to 1/2 of a turn. See
.
Figure 2-67 Loosening the Nut