INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 162

3 - 68
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
5
When powered on or taken out of Extended Standby, the HAPSITE will
automatically start preparing the default Headspace method. If the Headspace
method that the HAPSITE begins preparing is not the desired one, refer to
Section 3.2, Selecting a Different Method Using the SELECT METHOD Icon,
on page 3-13
.
6
The HAPSITE will display the progress of its preparation. Refer to
, Step 5 for more information.
6a
If running a Headspace method that requires a Concentrator. The HAPSITE will
run a Concentrator Cleanout. Its progress will be shown by a bar graph. If the
Concentrator Cleanout is not successful, refer to section
Concentrator Cleanout Failure, on page 3-11
.
7
When the HAPSITE has finished preparing, a SYSTEM READY message will
appear with a prompt to press Survey or Analyze to begin sampling.
NOTE: A blank run is recommended before running a sample.
8
Using the touch screen, touch RUN ANALYZE. See
Figure 3-112
. If using the
push buttons, push ANALYZE RUN.
Figure 3-112 HSS Analyze
9
Gently insert the Headspace needle into the septum of the 40 mL vial
containing the blank/sample.
10
The screen will prompt Collecting Sample For. See
Figure 3-113
. The
Headspace is automatically collecting a sample. No action is required by the
user.