INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 429
12 - 11
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
2
After the standard(s) have been prepared, a method must be developed to
collect the data. The method can be a default method or a modified method
using the Method Editor. All method development and chromatographic
changes which can affect the retention time must be made prior to collecting
the standard runs that are to be used for building and calibrating the library.
Refer to
for additional information on method
development.
HINT: When running each sample, enter the concentration and a description
on the Data File Information page. See
. To access this page
refer to
section 9.3, Data Review Toolbar, on page 9-5
Figure 12-9 Data File Information Page
NOTE: In this example, a series of standards at four concentration levels were
prepared by serial dilution. The method, QuantExample, was used to
collect the data. Start with one method for collection and continue with
that method. These are arbitrary concentrations for example only and
NOT actual concentrations.
QuantExample_20060301_005 = 150 PPM
QuantExample_20060301_004 = 50 PPM
QuantExample_20060301_002 = 200 PPM
QuantExample_20060301_001 =100 PPM
3
After the data has been collected, open the Calibration function which can be
accessed from the drop-down menu, by right mouse button clicking on the
HAPSITE icon, by the toolbar icon or via the Status icon and Function tab.
See
Figure 12-10
.