1 quick reference sop — gc/ms mode with hss – INFICON HAPSITE Smart Plus Chemical Identification System User Manual
Page 230

5 - 24
IP
N 07
4-
47
2-
P1
C
HAPSITE Smart Plus Operating Manual
5.6.1 Quick Reference SOP — GC/MS Mode with HSS
1
Double-click the Plus IQ software icon.
2
Verify that the desired sample configuration (i.e., Sample Loop or
Concentrator) is installed.
3
Power on or take the HAPSITE out of Extended Standby.
4
Depending upon the method, the HAPSITE may initiate a Concentrator
cleanout. If needed, see
Section 3.1.2, Concentrator Options (CONC
. See
Section 5.3, Selecting a New Method, on page
if the start up method is not the desired method.
5
Following a 20 minute sample temperature equilibration in the HSS oven,
gently press the Headspace needle into the 40 mL sample vial.
6
When the system has finished preparing, click the RUN METHOD icon and
select the desired method. Refer to Steps 2-4 of
7
Click the Run button in the pop-up window.
8
When the screen prompts Collecting Sample Now and Sampling Is Done, no
action is required from the user.
9
When the run is complete, a METHOD FINISHED prompt will appear.
10
When prompted, place the Headspace needle in the purge vial and press OK.
11
To analyze the data, see
.