Forcing Errors
All signals displayed in blue may be changed to force errors and
anomalies.
Parity Error
Parity error is checked by the recipient of data. If PAR is not in
compliance with data, PERR is asserted two clocks later. The TA700
behaves the same way, meaning that it can only assert PERR when it is
the intended recipient of data. Although it seems as if PERR is
programmable during all transactions it is only altered per the users
program during the data phase of Read transactions generated by the
TA700.
Wait States
Wait states may be inserted by pointing the cursor on the line where a
Wait State is desired and then clicking Insert.
Looping
You may define a loop by clicking in the Loop Start column next to the
program line where you wish for the loop to start and in the Loop End
column next to the program line where you wish for the loop to end. Loop
end may be placed in Idle lines.
Enable Loop
For the loop to execute, check the Autorun check box and enter a value
of up to 65535 in the Number of times edit box.