Input circuit monitor 277, Temperature compensation, Measured value correction – West Control Solutions KS98-1 User Manual
Page 277
Temperature compensation
Measurement of the cold-junction reference temperature is using a PTC resistor. The temperature error thus deter -
mined is converted into mV of the relevant thermocouple type, linearized and added to the measured value as correc -
tive value with correct polarity. The remaining error with varying cold-junction reference temperature is approx.
0,5K/10K, i.e. about one tenth of the error which would occur without compensation. Better results are possible
with controlled external TC, which is adjustable within 0...+140°C at the cold junction reference dependent of con -
trolled temperature. With cold-junction reference measurements for “reproducibility” assessment, however, utmost
care must be taken that constant environmental conditions are not exceeded when working with internal TC. An air
draft at the PTC resistor of the cold junction reference can be sufficient to falsify the measurement result.
Measured value correction
The measurement can be corrected in various ways using the measured value correction.
Pre-requisite: configuration
XKorr = ein. In most cases, the relative accuracy and reproducibility rather than the
absolute one are of interest, e.g.:
-the compensation of measurement errors in one working point ( control)
-minimization of linearity errors in a limited operating range (variable )
-correspondence with other measuring facilities (recorders, indicators, PLCs, ...)
-compensation of sample differences of sensors, transmitters, etc.
Measured value correction is designed for zero offset, gain matching and for both. It corresponds to scaling mx+b, with the
difference that the KS 98-1 firmware calculates gain m and zero offset b from the value pairs for process value (
x1in;
x2in) and (x1out; x2out) of two reference points.
Example 1:
Zero offset
x1in = 100°C
x1out = 100°C + 1,5°C
x2in = 300°C
x2out = 300°C + 1,5°C
The corrected values are shifted evenly with reference
to the input values over the complete range.
Example 2:
Gain change (rotation around the coordinate origin)
x1in =
0°C
x1out = 0°C
x2in = 300°C
x2out = 300°C + 1,5°C
The corrected values diverge despite equality with the
input values at
x1in
and
x1out
.
Inputs
9499-040-82711
III-277
AINP1 ( analog input 1 (No. 08))
X2out
X2in
X1out
X1in
korrigierte Kennlinie
Eingangs-Kennlinie
corrected characteristic
input characteristic
Fig. 98
X2out
X2in
X1out X1in
=
korrigierte Kennlinie
Eingangs-Kennlinie
corrected characteristic
input characteristic
Fig. 99