4.3 Measurement of Harmonic Current

Switching power sources are widely used in both consumer and industrial electrical products. The

capacitor input type rectifier circuits that are used with such input power lines have disadvantages such as

significant distortion of input current and a large amount of harmonic current. When a large amount of this

current flows in the power source line, the equipment may experience operation faults, such as voltage

distortion, which can cause the transformer to overheat, possibly resulting in a hazardous accident.

The APS-1102A includes a function for measuring harmonic current. This measurement does not

conform to the IEC or other standards.

This measurement can be performed only in AC-INT mode and when the fundamental wave (panel

frequency setting) is either 50 Hz or 60 Hz. Absolute values of harmonic current in RMS and their

percentages to the fundamental are displayed up to the 40th harmonics (2 kHz at 50 Hz fundamental).

The measurement value of harmonic current is reference value.

When harmonic current measurement is selected, RMS value, average DC, and peak value measurements

are not performed. If RMS value, average DC, or peak value has been set, harmonic current measurement

is not performed.

a) Changing to harmonic measurement window

To view the harmonic measurement window, select “MEASURE” in the MISC menu

.

Operation steps

1. Press the

or

key to move the cursor to the

icon, then press the

EN TER

key.

ENTER

CANCEL

Cursor

The MISC menu is displayed.

2. Press the

or

key to move the cursor to “MEASURE”, then press the

EN TER

key.

The selection box is displayed.

ENTER

CANCEL

Selection box