beautypg.com

GW Instek APS-1102A User Manual User Manual

Page 275

background image

9 SPECIFICATIONS

APS-1102A

9-9

Average value:

Average value of AC+DC (measures DC component)

Full scale:

15.00 A

Display resolution:

0.01 A

Measurement accuracy:

(|0.5% of rdg| + 0.08 A)

(at 23

5°C, output current is 5% to 100% of the maximum

current)

Peak value:

Minimum current and maximum current are displayed

separately

Full scale:

45.0 Apk

Display resolution:

0.1 Apk

Measurement accuracy:

(|2% of rdg| + 0.4 Apk)

(reference value, at 23

5°C, sine wave of which frequency is 45

Hz to 65 Hz)

Peak value hold:

Maximum value of

max current

and

min current |

Full scale:

45.0 Apk

Display resolution:

0.1 Apk

c) Output power measurement

Effective power

Full scale:

1200 W

Display resolution:

1 W

Measurement accuracy:

at 45 Hz to 65 Hz

(2% of rdg + 1 W)

at DC

(3% of rdg + 12 W)

(at 23

5°C, load with the power factor 0.5 to lower than 1, 50 V

or higher output voltage, output current is 10% to 100% of the

maximum current)

Apparent power:

Calculated as output voltage RMS value × output current RMS

value

Full scale:

1400 VA

Display resolution:

1 VA

Measurement accuracy:

at 45 Hz to 65 Hz

(2% of rdg + 1 VA)

at DC

(3% of rdg + 12 VA)

(at 23

5°C, 50 V or higher output voltage, output current is

10% to 100% of the maximum current)