GW Instek APS-1102A User Manual User Manual
Page 220

APS-1102A User Manual
APS-1102A
6-36
e) DISPlay subsystem
DISPlay[:WINDow]:MEASure:MODE
This selects the measurement display selection. The measurement display selection setting is
RMS (RMS value), AVG (average DC), PEAK (peak value), and HC (harmonic current) 1 to 4.
The relation between the harmonic current mode and the harmonic degree is as follows.
HC1 ( 1st to 10th)
HC2 (11th to 20th)
HC3 (21st to 30th)
HC4 (31st to 40th)
However, when in AC-INT mode, if the frequency is neither 50 Hz nor 60 Hz, the harmonic
current cannot be measured. ([3, “Invalid in this mode”] or [6, “Invalid in this frequency”] error
occurs.)
When the measurement display is set to HC1-HC4, commands except for
MEAS[:SCAL]:CURR:HARMonic[:AMPL]? (
See
“MEASure[:SCALar]:CURRent:HARMonic[:AMPLitude]?” on page 6-32.) or
MEAS[:SCAL]:CURR:HARM:RAT? (
See
“MEASure[:SCALar]:CURRent:HARMonic:RATio?” on page 6-32.) will not be accepted.
([4,"Under Busy State"] error will come out.)
Syntax:
DISPlay[:WINDow]:MEASure:MODE
Parameters:
RMS/AVG/PEAK/HC1/HC2/HC3/HC4
Command example:
DISP:MEAS:MODE RMS
Query example:
DISP:MEAS:MODE?
Response format:
Response example:
RMS