e) DISPlay subsystem

DISPlay[:WINDow]:MEASure:MODE

This selects the measurement display selection. The measurement display selection setting is

RMS (RMS value), AVG (average DC), PEAK (peak value), and HC (harmonic current) 1 to 4.

The relation between the harmonic current mode and the harmonic degree is as follows.

HC1 ( 1st to 10th)

HC2 (11th to 20th)

HC3 (21st to 30th)

HC4 (31st to 40th)

However, when in AC-INT mode, if the frequency is neither 50 Hz nor 60 Hz, the harmonic

current cannot be measured. ([3, “Invalid in this mode”] or [6, “Invalid in this frequency”] error

occurs.)

When the measurement display is set to HC1-HC4, commands except for

MEAS[:SCAL]:CURR:HARMonic[:AMPL]? (

See

MEASure[:SCALar]:CURRent:HARMonic[:AMPLitude]?” on page 6-32.) or

MEAS[:SCAL]:CURR:HARM:RAT? (

See

MEASure[:SCALar]:CURRent:HARMonic:RATio?” on page 6-32.) will not be accepted.

([4,"Under Busy State"] error will come out.)

Syntax:

DISPlay[:WINDow]:MEASure:MODE

Parameters:

RMS/AVG/PEAK/HC1/HC2/HC3/HC4

Command example:

DISP:MEAS:MODE RMS

Query example:

DISP:MEAS:MODE?

Response format:

Response example:

RMS