4 setting output frequency, 5 output initial phase settings – GW Instek APS-1102A User Manual User Manual
Page 153
5 MENUS
APS-1102A
5-19
5.3.4 Setting output frequency
This sets the AC voltage frequency.
When either AC-SYNC or AC+DC-SYNC mode is selected, the AC voltage frequency cannot be set.
Table5-14. Output Frequency Setting
Setting Range
Resolution
Setting Accuracy
1.0 Hz to 550.0 Hz
0.1 Hz
0.01% (1.0 Hz to 550.0 Hz,23
5
C)
S
ee “3.4.7 Using limiter functions”, when using a setting range limit for output frequency.
Operation steps
To set the output frequency, set a numerical value labeled
in the SET menu.
See
“3.4.6 Setting output frequency”, for description of output frequency settings.
S
ee “5.5.5 Setting range limiter (LIMIT)”, for description of setting range limits.
-------
!
CAUTION ----------------------------------------------------------------------------------------------
If the frequency is set under 40 Hz in AC-INT mode, there is a possibility that the output
waveform will be distorted and DC offset will be output.
----------------------------------------------------------------------------------
5.3.5 Output initial phase settings
When AC-INT, AC-ADD, AC-SYNC, AC+DC-INT, AC+DC-ADD, or AC+DC-SYNC mode is selected,
the AC voltage phase can be set according to the internal signal source when output is on.
Table5-15. Output Initial Phase Settings
Setting Range
Resolution
0.0
to 359.9
0.1
Operation steps
To set the phase, select
in the SET menu and set a numerical value.
For description of the output initial phase:
See