beautypg.com

4 setting output frequency, 5 output initial phase settings – GW Instek APS-1102A User Manual User Manual

Page 153

background image

5 MENUS

APS-1102A

5-19

5.3.4 Setting output frequency

This sets the AC voltage frequency.

When either AC-SYNC or AC+DC-SYNC mode is selected, the AC voltage frequency cannot be set.

Table5-14. Output Frequency Setting

Setting Range

Resolution

Setting Accuracy

1.0 Hz to 550.0 Hz

0.1 Hz

0.01% (1.0 Hz to 550.0 Hz,23

5

C)

S

ee “3.4.7 Using limiter functions”, when using a setting range limit for output frequency.

Operation steps

To set the output frequency, set a numerical value labeled

in the SET menu.

See

“3.4.6 Setting output frequency”, for description of output frequency settings.

S

ee “5.5.5 Setting range limiter (LIMIT)”, for description of setting range limits.

-------

CAUTION ----------------------------------------------------------------------------------------------

If the frequency is set under 40 Hz in AC-INT mode, there is a possibility that the output

waveform will be distorted and DC offset will be output.

----------------------------------------------------------------------------------

5.3.5 Output initial phase settings

When AC-INT, AC-ADD, AC-SYNC, AC+DC-INT, AC+DC-ADD, or AC+DC-SYNC mode is selected,

the AC voltage phase can be set according to the internal signal source when output is on.

Table5-15. Output Initial Phase Settings

Setting Range

Resolution

0.0

to 359.9

0.1

Operation steps

To set the phase, select

in the SET menu and set a numerical value.

For description of the output initial phase:

See

“4.2.2 Set phase at output on”.