Protection function-related errors, Panel operation errors, 2 protection function-related errors – GW Instek APS-1102A User Manual User Manual
Page 248

APS-1102A User Manual
APS-1102A
7-6
7.2.2 Protection function-related errors
When a protection function-related error has occurred, an error message is displayed in the message
window, and operation is stopped.
Figure 7-2. Screen Display When Protection Function-Related Error Has Occurred
When a protection function-related error message is displayed, turn off the power switch. (
See “7.1
Protection Function
”.)
Table7-3 lists error messages and actions when errors are detected.
Table7-3. Protection Function-Related Errors
Error Message
Protection Item
Cause and Operation or Action Required
When Fault Is Detected
System Locked.
caused by:
COMMUNICATION FAILURE 1
Internal
communication
error 1
Output is set to off and system lock status is
set. Turn off the power switch.
If the same protection error occurs again after
the power is turned back on, contact GW
Instek or our agent.
System Locked.
caused by:
COMMUNICATION FAILURE 2
Internal
communication
error 2
System Locked.
caused by:
FRONT PANEL FAILURE
Control panel
error
System Locked.
caused by:
AMP NO RESPONSE ERROR
AMP no response
error
System Locked.
caused by:
FAN FAILURE ERROR
Fan failure error