beautypg.com

5 remote-related external control errors – GW Instek APS-1102A User Manual User Manual

Page 253

background image

7 TROUBLESHOOTING

APS-1102A

7-11

7.2.5 Remote-related external control errors

Table7-8 lists remote-related external control error messages.

Table7-8. Remote Error Message List

No.

Message Text

Cause

0

“No error”

No error.

100 “Command error”

Command was recognized as invalid due to any cause.

(General command error)

102 “Syntax error”

Command was recognized as invalid due to reception of
undefined command or parameter.

200 “Execution error”

Error occurred when executing the command due to any cause.
(General execution error)

222 “Data out of range”

Error occurred when executing the command due to out of range
parameter.

350 “Too Many Errors”

New errors cannot be retained due to error queue overflow.

1

“Invalid with output on”

Error occurred when executing the command because the output
relay was on.

2

“Invalid with output off”

Error occurred when executing the command because the output
relay was off.

3

“Invalid in this mode”

Error occurred when executing the command due to inappropriate
operation mode.

4

“Under Busy State”

Error occurred during execution of the command, because the
device is busy state.

5

“Out of Limiter”

Error occurred during execution of the command because limiter
value was exceeded.

6

“Invalid in this frequency”

Error occurred when executing the command due to inappropriate
oscillation frequency setting.

7

“Out of Input Signal
frequency”

Error occurred when executing the command due to out of the
external synchronization frequency range.

8

“Under Warning State”

Error occurred when executing the command because device was
operating in protected mode.