5-24
5.4.1 Output voltage measurement
When “MEASURE” has been selected in the MISC menu, RMS value (RMS), average DC (AVG), or
peak value (PEAK) is displayed.
The initial setting (factory setting) is to display the RMS value (RMS).
See “3.4.9 Using measurement functions”.
a) RMS value
RMS value of AC+DC
Full scale:
250.0 Vrms: 100 V range
500.0 Vrms: 200 V range
Display resolution:
0.1 Vrms
Measurement accuracy: at 45 Hz to 65 Hz
(0.5% of rdg + 0.3 Vrms): 100 V range
(0.5% of rdg + 0.6 Vrms): 200 V range
at DC, 40 Hz to 550 Hz
(0.7% of rdg + 0.9 Vrms): 100 V range
(0.7% of rdg + 1.8 Vrms): 200 V range
(at 23
5°C)
b) Average value
Average value of AC+DC (measurement of DC)
Full scale:
250.0 V: 100 V range
500.0 V: 200 V range
Display resolution:
0.1 V
Measurement accuracy:
(|0.5% of rdg| + 0.4 V): 100 V range
(|0.5% of rdg| + 0.8 V): 200 V range
(at 23
5°C)
c) Peak value
Minimum voltage and maximum voltage are displayed
separately
Full scale:
250 Vpk: 100 V range
500 Vpk: 200 V range
Display resolution:
1 Vpk
Measurement accuracy:
(|1.5% of rdg| + 3 Vpk): 100 V range
(|1.5% of rdg| + 6 Vpk): 200 V range
(reference value, at 23
5°C, sine wave of which frequency is 45 Hz to 65
Hz)