beautypg.com

Using limiter functions, 7 using limiter functions – GW Instek APS-1102A User Manual User Manual

Page 62

background image

APS-1102A User Manual

APS-1102A

3-28

3.4.7 Using limiter functions

a) Using current limiter

The APS-1102A includes a peak current limiter function and a RMS current limiter function.

Limiter settings are retained separately for 100 V and 200 V ranges.

Setting range limits can be set for output voltage and output frequency.

See “5.3.7 Set Peak current limiter settings”.

See “5.3.8 Set RMS current limiter settings”.

Table3-12. Current Limiter Setting Ranges

Setting

Setting Range

Resolution

Initial Value

Peak current
limiter (positive)

100 V range

+10.0 A to +42.0 A

0.1 A

+42.0 A

200 V range

+5.0 A to +21.0 A

0.1 A

+21.0 A

Peak current
limiter (negative)

100 V range

42.0 A to

10.0 A

0.1 A

42.0 A

200 V range

21.0 A to

5.0 A

0.1 A

21.0 A

RMS current
limiter

100 V range

1.0 A to 10.5 A

0.1 A

10.5 A

200 V range

1.0 A to 5.3 A

0.1 A

5.3 A

For a capacitive load or inductive load, the peak current limiter may not achieve the peak current

value as set, due to overshooting or undershooting the set current value, or variable output current

waveforms while the limiter is operating.

For an inductive load, operation of the limiter can cause the current to vary and, as a result, an output

overvoltage may occur.

Note with caution that the peak current limiter may allow overshooting, depending on the load, and

therefore may not achieve the peak current value as set.

For an inductive load, operation of the limiter can cause an output overvoltage, triggering the

overvoltage protection function. In this case, the peak current limiter settings (positive and

negative) should be reduced.

If the output current or frequency suddenly increases, due to an output short, etc., the limiter

response may not be quick enough, so the output overcurrent protection function may be triggered.

CAUTION

CAUTION