GW Instek APS-1102A User Manual User Manual
Page 219

6 REMOTE INTERFACE
APS-1102A
6-35
MEASure[:SCALar]:VOLTage:HIGH?
This returns the maximum peak value (AC+DC peak value) of the output voltage in Vpk units. For
the 100 V range, the measurement full scale of the maximum peak value of output voltage is ±250
Vpk, and for the 200 V range, it is ±500 Vpk. If the measured value exceeds the measurement
range, 999 Vpk is returned as a measurement error.
Syntax:
MEASure[:SCALar]:VOLTage:HIGH?
Parameters:
None
Query example:
MEAS:VOLT:HIGH?
Response format:
Response example:
100
MEASure[:SCALar]:VOLTage:LOW?
This returns the minimum peak value (AC+DC peak value) of the output voltage in Vpk units. For
the 100 V range, the measurement full scale of the minimum peak value of output voltage is ±250
Vpk, and for the 200 V range, it is ±500 Vpk. If the measured value exceeds the measurement
range,
999 Vpk is returned as a measurement error.
Syntax:
MEASure[:SCALar]:VOLTage:LOW?
Parameters:
None
Query example:
MEAS:VOLT:LOW?
Response format:
Response example:
99
MEASure[:SCALar]:VOLTage:AVErage?
This returns the output voltage average value (AC+DC average value) in V units. For the 100 V
range, the measurement full scale of the average value of output voltage is ±250.0 V, and for the
200 V range, it is ±500.0 V. If the measured value exceeds the measurement range, 999.9 V is
returned as a measurement error.
Syntax:
MEASure[:SCALar]:VOLTage:AVErage?
Parameters:
None
Query example:
MEAS:VOLT:AVE?
Response format:
Response example:
150.2