GW Instek APS-1102A User Manual User Manual
Page 215
6 REMOTE INTERFACE
APS-1102A
6-31
MEASure[:SCALar]:CURRent:AMPLitude:MAX?
This returns the output current peak value (hold) in Apk units. The held output current peak value
full scale is 45.0 Apk. If the measured value exceeds the measurement range, 99.9 Apk is returned
as a measurement error.
Syntax:
MEASure[:SCALar]:CURRent:AMPLitude:MAX?
Parameters:
None
Query example:
MEAS:CURR:AMPL:MAX?
Response format:
Response example:
26.5
MEASure[:SCALar]:CURRent:AMPLitude:RESet
This resets the stored output current peak value (hold) to 0 Apk.
Syntax:
MEASure[:SCALar]:CURRent:AMPLitude:RESet
Parameters:
1
Command example:
MEAS:CURR:AMPL:RES 1
Response format:
None
Response example:
None
MEASure[:SCALar]:CURRent:AVErage?
This returns the output current average value (AC+DC average value) in a units. The output
current average value measurement full scale is
15.00 A. If the measured value exceeds the
measurement range, 99.99 A is returned as a measurement error.
Syntax:
MEASure[:SCALar]:CURRent:AVErage?
Parameters:
None
Query example:
MEAS:CURR:AMPL?
Response format:
Response example:
1.23
MEASure[:SCALar]:CURRent:CREStfactor?
This returns the output current crest factor. The crest factor measurement full scale is 50.00. If the
measured value exceeds the measurement range, 99.99 is returned as a measurement error.
The crest factor is calculated as “current peak value/current RMS value”.
Syntax:
MEASure[:SCALar]:CURRent:CREStfactor?
Parameters:
None
Query example:
MEAS:CURR:CRES?
Response format:
Response example:
1.41