6-34

MEASure[:SCALar]:POWer:AC:REACtive?

This returns the reactive power in var units. The reactive power measurement full scale is 1400 var.

If the measured value exceeds the measurement range, 9999 var is returned as a measurement

error.

Syntax:

MEASure[:SCALar]:POWer:AC:REACtive?

Parameters:

None

Query example:

MEAS:POW:AC:REAC?

Response format:

Response example:

269

MEASure[:SCALar]:POWer:AC[:REAL]?

The effective power is returned in W units. The effective power measurement full scale is 1200 W.

If the measured value exceeds the measurement range, 9999 W is returned as a measurement error.

Syntax:

MEASure[:SCALar]:POWer:AC[:REAL]?

Parameters:

None

Query example:

MEAS:POW:AC?

Response format:

Response example:

249

MEASure[:SCALar]:VOLTage[:AC]?

This returns the output voltage RMS value (AC+DC RMS value) in Vrms units.

For the 100 V

range, the output voltage RMS value measurement full scale is 250.0 Vrms, and for the 200 V

range, it is 500.0 Vrms. If the measured value exceeds the measurement range, 999.9 Vrms is

returned as a measurement error.

Syntax:

MEASure[:SCALar]:VOLTage[:AC]?

Parameters:

None

Query example:

MEAS:VOLT?

Response format:

Response example:

100.4