Output, 1 output – GW Instek APS-1102A User Manual User Manual
Page 268

APS-1102A User Manual
APS-1102A
9-2
Unless otherwise noted, the following conditions are adopted:
Load:
Resistance load of power factor 1
Output mode:
INT (internal signal source)
Output waveform:
Sine wave
Current Limiter:
Factory default setting
Output terminals:
Terminal block on rear panel
[set] indicates a setting value, and [rdg] indicates a reading value.
9.1 Output
a) Output mode
All eight modes (combination of operation mode + signal source mode)
AC-INT mode
(alternate current - internal signal source)
AC-EXT mode
(alternate current - external signal source)
AC-ADD mode
(alternate current - internal + external source)
AC-SYNC mode
(alternate current - external synchronization)
AC+DC-INT mode
(direct current - internal signal source)
AC+DC-EXT mode
(direct current - external signal source)
AC+DC-ADD mode
(direct current - internal + external source)
AC+DC-SYNC mode
(direct current - external synchronization)
b) Output voltage ranges
100 V range and 200 V range
c) Maximum output power
750 VA (AC) / 750 W (DC)
Power supply input conditions:
100 V AC to 180 V AC input
(hereinafter referred to as “AC 100 V input system”)
1000 VA (AC) / 1000 W (DC)
Power supply input conditions:
180 V AC to 250 V AC input
(hereinafter referred to as “AC 200 V input system”)
d) Load power factor
0 to 1 (phase lead or phase lag)
Note: External power injection and regeneration are not
available.