2.3 Grounding and Power Supply Connection
Be sure to ground the APS-1102A.
This product uses a line filter. Be sure to ground this product. Otherwise, an electric shock
may occur.
To prevent an electric shock from occurring, be sure to ground the APS-1102A according to
"Electric Equipment Technical Standard Class D (100
or less) Grounding" or higher.
When the input power line is 100 V AC, use power cord set 1. When a 3-prong power plug that
includes a protective ground contact is connected to a 3-prong power supply outlet, this product is
grounded automatically. Power cord set 1 (Japan version) is rated at 125 V AC.
When the input power line is 200 V AC, use power cord set 2 equipped with a terminal or crimp
contact that suits the type of outlet to be used. Always connect a ground when using power cord set 2.
Power cord set 2 is rated at 250 V AC.
A 3-prong/2-prong conversion adapter is not provided with this product. If a 3-prong/2-prong
conversion adapter is used, the adaptor ground wire must be connected to a ground contact next to the
outlet.
The power code set can be used for disconnecting the product from AC power line in case of
emergency. Maintain enough space around the inlet, to be able to remove the connector of a power
cord from the inlet. Use a power socket located at convenient place with adequate space around so
that the plug can be removed from socket
.
The power requirements of this product are as follows:
Voltage range:
100 V AC to 230 V AC
10 % (when at 250 V or less)
Overvoltage category II
Frequency range:
50 Hz
2 Hz /60 Hz
2 Hz (single phase)
Power consumption:
1.4 kVA or less
!