APS-1102A User Manual

APS-1102A

6-30

d) MEASure subsystem

MEASure[:SCALar]:CURRent[:AC]?

This returns the output current RMS value (AC+DC RMS value) in Arms units. The output current

RMS value measurement full scale is 15.00 Arms. If the measured value exceeds the measurement

range, 99.99 Arms is returned as a measurement error.

Syntax:

MEASure[:SCALar]:CURRent[:AC]?

Parameters:

None

Query example:

MEAS:CURR?

Response format:

Response example:

12.85

MEASure[:SCALar]:CURRent:HIGH?

This returns the highest current peak value (AC+DC peak value) in Apk units. The highest current

peak value measurement full scale is

45.0 Apk. If the measured value exceeds the measurement

range, 99.9 Apk is returned as a measurement error.

Syntax:

MEASure[:SCALar]:CURRent:HIGH?

Parameters:

None

Query example:

MEAS:CURR:HIGH?

Response format:

Response example:

25.9

MEASure[:SCALar]:CURRent:LOW?

This returns the lowest current peak value (AC+DC peak value) in Apk units. The lowest current

peak value measurement full scale is

45.0 Apk. If the measured value exceeds the measurement

range,

99.9 Apk is returned as a measurement error.

Syntax:

MEASure[:SCALar]:CURRent:LOW?

Parameters:

None

Query example:

MEAS:CURR:LOW?

Response format:

Response example:

35.7