3.4.3 Setting output voltage range

Select 100 V range or 200 V range. The output voltage range cannot be changed while output is on.

The setting range varies according to the output voltage range.

Table3-8. List of Setting Ranges for Various Output Voltage Range

Setting

Setting Range

100 V range

200 V range

Unit

DC voltage

220.0 to +220.0

440.0 to +440.0

V

AC voltage

Wave-
form

SIN/SQU

0.0 to 155.0

0.0 to 310.0

Vrms

ARB1 to 16

0.0 to 440.0

0.0 to 880.0

Vp-p

Peak current limiter (positive)

+10.0 to +42.0

+5.0 to +21.0

A

Peak current limiter (negative)

42.0 to

10.0

21.0 to

5.0

A

RMS current limiter

1.0 to 10.5

1.0 to 5.3

A

Positive voltage setting limit

+0.1 to +220.0

+0.1 to +440.0

V

Negative voltage setting limit

220.0 to

0.1

440.0 to

0.1

V

External input gain

0.0 to 220.0

0.0 to 440.0

times

Operation steps

To set the output voltage range, select

in the SET menu.

1. Press the

or

key to move the cursor to the

icon.

ENTER

CANCEL

Cursor