Current limiter, Setting range limits, Signal sources – GW Instek APS-1102A User Manual User Manual
Page 272: 2 current limiter, 3 setting range limits

APS-1102A User Manual
APS-1102A
9-6
9.2 Current Limiter
a) Output peak current limiter When limiter is operating, output voltage is clipped.
Positive current setting range:
100 V range: +10.0 A to +42.0 A (Initial value: +42.0 A)
200 V range: +5.0 A to +21.0 A (Initial value: +21.0 A)
Negative current setting range:
100 V range:
42.0 A to
10.0 A (Initial value:
42.0 A)
200 V range:
21.0 A to
5.0 A (Initial value:
21.0 A)
Setting resolution:
0.1 A
b) Output RMS current limiter When limiter is operating, suppresses output voltage.
Current setting range:
100 V range:
1.0 A to 10.5 A (Initial value: 10.5 A)
200 V range:
1.0 A to
5.3 A (Initial value: 5.3 A)
Setting resolution:
0.1 A
9.3 Setting Range Limits
When signal source mode is INT or ADD only
a) Voltage setting limit
Positive voltage setting range:
100 V range: +0.1 V to +220.0 V (Initial value: +220.0 V)
200 V range: +0.1 V to +440.0 V (Initial value: +440.0 V)
Negative voltage setting range:
100 V range:
220.0 V to
0.1 V (Initial value:
220.0 V)
200 V range:
440.0 V to
0.1 V (Initial value:
440.0 V)
Setting resolution:
0.1 V
b) Frequency setting limit
(Lower limit
upper limit)
Upper limit setting range:
1.0 Hz to 550.0 Hz
(Initial value: 550.0 Hz)
Lower limit setting range:
1.0 Hz to 550.0 Hz
(Initial value: 1.0 Hz)
Setting resolution:
0.1 Hz