beautypg.com

Rms current limiter settings, Measure screen, 8 rms current limiter settings – GW Instek APS-1102A User Manual User Manual

Page 156

background image

APS-1102A User Manual

APS-1102A

5-22

5.3.8 RMS current limiter settings

This limits the output RMS current.

When the output RMS current exceeds this setting, the output voltage is suppressed.

Table5-17. RMS Current Limiter Setting Range

Setting

Setting Range

Resolution

Initial Value

RMS current
limiter

100 V range

1.0 A to 10.5 A

0.1 A

10.5 A

200 V range

1.0 A to 5.3 A

0.1 A

5.3 A

Operation steps

Select

in the SET menu and set numerical values.

See

“3.4.7 Using limiter functions”, for description of the output average value current

limiter.