5.4.3 Output power measurement

The output power values described below are always displayed.

a) Effective power

Full scale:

1200 W

Display resolution:

1 W

Measurement accuracy: at 45 Hz to 65 Hz

(2% of rdg +1 W)

at DC

(3% of rdg +12 W)

(at 23

5°C, load with the power factor 0.5 to lower than 1, 50 V or higher

output voltage, output current is 10% to 100% of the maximum current)

b) Apparent power

Calculated as output voltage RMS value × output current RMS value

Full scale:

1400 VA

Display resolution:

1 VA

Measurement accuracy: at 45 Hz to 65 Hz

(2% of rdg +1 VA)

at DC

(3% of rdg +12 VA)

(at 23

5°C, 50 V or higher output voltage, output current is 10% to 100%

of the maximum current)

c) Reactive power

Calculated as

(

)

apparent power 2-

(

)

effective power 2

Full scale:

1400 var

Display resolution:

1 var

Measurement accuracy:

(2% of rdg +1 var)

(at 23

5°C, load with the power factor 0 to lower than 0.5, 45 Hz to 65 Hz,

50 V or higher output voltage, output current is 10% to 100% of the

maximum current)

5.4.4 Load power factor measurement

Calculated by effective power/apparent power

In the case that it exceeds the measurement range, “



” is displayed.

Measurement range:

0.00 to 1.00

Display resolution:

0.01