Output power measurement, Load power factor measurement, Load crest factor measurement – GW Instek APS-1102A User Manual User Manual
Page 160: Output harmonic current measurement, External synchronization frequency measurement, 3 output power measurement, 4 load power factor measurement

APS-1102A User Manual
APS-1102A
5-26
5.4.3 Output power measurement
The output power values described below are always displayed.
a) Effective power
Full scale:
1200 W
Display resolution:
1 W
Measurement accuracy: at 45 Hz to 65 Hz
(2% of rdg +1 W)
at DC
(3% of rdg +12 W)
(at 23
5°C, load with the power factor 0.5 to lower than 1, 50 V or higher
output voltage, output current is 10% to 100% of the maximum current)
b) Apparent power
Calculated as output voltage RMS value × output current RMS value
Full scale:
1400 VA
Display resolution:
1 VA
Measurement accuracy: at 45 Hz to 65 Hz
(2% of rdg +1 VA)
at DC
(3% of rdg +12 VA)
(at 23
5°C, 50 V or higher output voltage, output current is 10% to 100%
of the maximum current)
c) Reactive power
Calculated as
(
)
apparent power 2-
(
)
effective power 2
Full scale:
1400 var
Display resolution:
1 var
Measurement accuracy:
(2% of rdg +1 var)
(at 23
5°C, load with the power factor 0 to lower than 0.5, 45 Hz to 65 Hz,
50 V or higher output voltage, output current is 10% to 100% of the
maximum current)
5.4.4 Load power factor measurement
Calculated by effective power/apparent power
In the case that it exceeds the measurement range, “
” is displayed.
Measurement range:
0.00 to 1.00
Display resolution:
0.01