2 PREPARATIONS BEFORE USE

APS-1102A

2-9

2. Panel settings

Set the APS-1102A as shown below.

See “3.4 Main Operation Example”, for descriptors

of the settings methods.

Table2-1. Panel Settings for Operation Check

Item

Setting

Output mode

AC-INT mode

Output voltage range

100 V

AC output voltage

100 Vrms

Selected measurement display RMS

3. Press the

OUTPUT

key to turn on output.

OUTPUT

OUTPUT

Output is off.

Output is on.

⇒ Check that the control panel MEASURE screen shows 100 Vrms as the AC output voltage.

Figure 2-3. Operation Check

⇒ If the normal values are not displayed, it may be due to an operation fault. Contact GW Instek or

our agent.

4. When you turn off the power, check the OUTPUT on/off LED is off. Then press the power

switch to the bottom () position. This shuts off the power supply and turns off the power

source unit.

Before turning off the power, make sure that the output on/off LED is off.

Output on/off LED is on.

Check display.

CAUTION