Table4-10. Setting Items in PROGRAM Screen (3/3)

Setting Name

Item

Operation

STEP TERM

Step end

Sets operation after end of step.
Select one of the following three options.

CONT:

Sequence continuation

STOP:

Change to Idle mode

HOLD:

Change to Hold mode

JUMP STEP

Jump step

Sets the step number to jump to after the end of the
previous step.

0:

Continue to next step number

1 to 255: Jump to specified step number

JUMP NUM

Jump times

Sets the number of times to execute the jump operation
specified by JUMP STEP.

0:

Infinite times jump

1 to 999: Specified times jump

BRANCH0

Branch step 0

Sets the step number to jump to when trigger input has
occurred.
Two types of branch steps can be set: BRANCH0 and
BRANCH1.
[Command example]

When a STOP is set after a small number of sweeps at the
jump destination step, it enables uses where there is a
safe region for saving data if necessary.

0:

Branch disabled

1 to 255: Jump to specified step number

BRANCH1

Branch step 1

The operation steps for each of the setting items in the PROGRAM screen are described below.

Operation step <1>

The settings for step number (STEP), step time (TIME), DC voltage (DC VOLT, AC+DC mode only),

AC voltage (AC VOLT), AC frequency (FREQ), phase (PHS), step end phase (STOP PHS), jump step

(JUMP STEP), jump times (JUMP NUM), and branch (BRANCH0/BRANCH1) are made by entering

numerical values for various items in the order described below.

The example described below is when entering the setting for step time (TIME).