7.3 When Failure Is Suspected

If a problem that might be due to a fault occurs when using the APS-1102A, check the following table to

determine whether the problem might be due to an operation, use method, connection error, etc.

If the problem is not covered by this table, leave the power off and contact GW Instek or our agent.

Table7-9. When Failure Is Suspected (Problem when Switching Power on/off)

Symptom

Possible Cause

Required Action, etc.

Does not operate after
turning on power
switch.

Not connected to a commercial
power source.

Make sure the power cord is firmly inserted into
the outlet and the inlet (on rear panel of the
APS-1102A).

Power source being used is
outside the rated range.

Use a power source that is within the rated range.

Internal fuse is broken.

Contact GW Instek or our agent.

Does not stop
immediately after
turning off power
switch.

This is not a failure.

Please wait. It will stop in approximately five
seconds.

LCD is blank.

(Fan is operating.)

LCD has low contrast setting.

Increase the LCD contrast setting.

To adjust the LCD contrast when the screen
contents cannot be seen, press the ENTER key
and CANCEL key at the same time. This sets the
digit cursor to the 10’s column.

See “5.6.3 LCD contrast

adjustment

.

Table7-10. When Failure Is Suspected (Problem During Key Operation)

Symptom

Possible Cause

Required Action, etc.

Panel operation does
not work.

Keylock has been set to on.

Turn off the keylock.

See “5.6.1 Keylock”.

Key or MODIFY dial has been
degraded.

Contact GW Instek or our agent for repairs.