beautypg.com

Sequence function – GW Instek APS-1102A User Manual User Manual

Page 276

background image

APS-1102A User Manual

APS-1102A

9-10

Reactive power:

Calculated as

(

)

apparent power 2-

(

)

effective power 2

Full scale:

1400 var

Display resolution:

1 var

Measurement accuracy:

(2% of rdg + 1 var)

(at 23

5°C, load with the power factor 0 to lower than 0.5, 45

Hz to 65 Hz, 50 V or higher output voltage, output current is

10% to 100% of the maximum current)

d) Load power factor measurement: Calculated as effective power / apparent power

Measurement range:

0.00 to 1.00

Display resolution:

0.01

e) Load crest factor measurement: Calculated as (maximum value of

maximum current

and

minimum current

) / RMS current

Measurement range:

1.00 to 50.00

Display resolution:

0.01

f) Output harmonic current measurement (AC - INT mode only, fundamental wave is 50 Hz

or 60 Hz only)

Measurement range:

To 40th harmonic of fundamental wave

Full scale:

15.00 Arms and 500.0%

Display resolution:

0.01 Arms and 0.1%

Measurement accuracy:

at RMS value current up to 10th

(1% of rdg + 0.08 Arms)

at RMS value current from 11th up to 20th

(2% of rdg + 0.08 Arms)

at RMS value current from 21st up to 40th

(5% of rdg + 0.08 Arms)

(All in the case of reference value, at 23

5°C)

Note: This measurement does not conform to the IEC or other

standards.

g) External synchronization frequency measurement (SYNC mode only)

Measurement range:

38.0 Hz to 525.0 Hz

Display resolution:

0.1 Hz

Measurement accuracy:

0.2 Hz (at 23

5°C)