beautypg.com

5 setting range limiter (limit) – GW Instek APS-1102A User Manual User Manual

Page 169

background image

5 MENUS

APS-1102A

5-35

5.5.5 Setting range limiter (LIMIT)

The setting range limiter indicates the LIMIT screen, enabling the setting range of output voltage and

output frequency to be limited when the signal source mode is the internal signal source mode (AC-INT,

AC+DC-INT) or internal + external signal source mode (AC-ADD, AC+DC-ADD), and the setting range

of the output voltage to be limited when in the external sync mode (AC-SYNC, AC+DC+SYNC).

The numeric values shown in Table5-23 are set on this screen.

See b) Using s

etting range limit in “3.4.7 Using limiter functions”.

Table5-23. Setting Range Limit (LIMIT)

Setting

Name

Setting

Output

Voltage

Range

Setting Range

Resolution

Initial

value

Unit

+Vo

Positive voltage
setting limit

100 V

+0.1 to +220.0

0.1

+220.0

V

200 V

+0.1 to +440.0

0.1

+440.0

V

Vo

Negative voltage
setting limit

100 V

220.0 to

0.1

0.1

220.0

V

200 V

440.0 to

0.1

0.1

440.0

V

F Up

Frequency setting upper limit

1.0 to 550.0

0.1

550.0

Hz

F

Lo

Frequency setting lower limit

1.0 to 550.0

0.1

1.0

Hz

 The setting range limit sets the limit for AC peak value + DC.

For example, when +Vo limit is +200.0 V, and the DC setting has already been set to 100.0 V, the

AC limit setting becomes 70.7 Vrms (100.0 Vpk).

 The voltage setting range limit that does not include an output voltage already set cannot be set.

 The same setting range limit is used for all output modes. Not only the currently selected output

mode setting, but the settings for all other modes are restricted by this limit. (The APS-1102A

stores the setting for each output mode.)

The following describes an example where the selected output mode is AC -INT and the limit is

set to prevent the output voltage from exceeding 100.0 Vrms.

In this case, the output limit range must be set to

141.4 V (100.0 Vrms), but this setting cannot

be made unless the output voltage setting is 100.0 Vrms or less for the other output modes where

the output voltage can be set, i.e., the AC-ADD, AC-SYNC, AC+DC-INT, AC+DC-ADD, and

AC+DC-SYNC modes. If the setting for AC-ADD is 110.0 Vrms, the setting range limit cannot

be set within

155.5 V (110.0 Vrms), even if the setting for the selected AC-INT is 100.0 Vrms.

Memory and sequences are not subject to setting range limits.