EXFO IQS-8100 Series Transport Blazer for IQS-600 User Manual

Page 97

background image

Creating and Starting a Test Case

SONET/SDH Application

83

Typical Test Cases

Creating an Optical SONET/SDH Test Case
(IQS-8115/20/30)

The following procedure describes a normal optical SONET/SDH the test
case on the IQS-8115/20/30 modules.

³

For Next generation test case, see Creating a Next Generation Test Case
including VCAT/LCAS and GFP
on page 103.

³

For IQS-8140, see Creating an Optical SONET/SDH Test Case on an
IQS-8140
on page 87.

To create an Optical SONET/SDH Test on an IQS-8115/20/30 module:
1.
Test configuration:

1a. Select the source Clock Mode that will be used

for the test. Refer to Clock Configuration on
page 129 for more information.

1b. Select Normal as the Test Mode. Refer to Test

Configuration on page 126 for more information.

1c. Select the Coupled check box to set the same

settings for both the TX and RX signals or clear
the Coupled check box to configure the TX and RX signal
individually (decoupled).

1d. Select the Through check box to loop the RX signal to the TX port.

The Clock Mode is automatically set to Recovered when the
Through check box is selected.

1e. Select the SONET/SDH Intrusive check box to loop the RX signal

to the TX port with TX overwrite capabilities. SONET/SDH
Intrusive
is not available on IQS-8115.

1f. Leave the OTN Intrusive check box cleared. OTN Intrusive is not

available on IQS-8115.

1g. Press Next.