EXFO IQS-8100 Series Transport Blazer for IQS-600 User Manual
Page 121
Creating and Starting a Test Case
SONET/SDH Application
107
Typical Test Cases
5. Select the payload: GFP or Pattern.
5a. For GFP, select the UPI, EXI, and CID
parameters. GFP is available in Coupled test
mode only. Refer to GFP Frame TX on page 432
for UPI and EXI, and to GFP Channel TX on
page 435 for CID. Press Next or Finish.
5b. For Pattern, go to step 7.
6. Select the GFP payload: Pattern or External Ethernet.
6a. For External Ethernet, which is available with
GFP, select the interface and its rate. Refer to
GFP Client TX on page 444 for more information.
Go to step 8.
6b. For Pattern, go to step 7.
7. Pattern Configuration:
7a. Set the pattern parameters. Refer to Pattern TX
on page 405 and Pattern RX on page 409 for
more information.
8. Press Finish to complete the test setup. The Grid tab
closes and automatically switch to the Alarm summary tab.
9. For additional configuration parameters and results, refer to Summary
Tabs on page 125, Port Tabs on page 137, OTN Tabs on page 153,
SONET Tabs on page 211, SDH Tabs on page 303, BERT Tabs on
page 405, Advanced Tabs on page 411, and Common Tabs on
page 491.
10. Press the Start button to start the test. Refer to Global Test Status and
Controls on page 35 for more information on test management.