background image

Creating and Starting a Test Case

90

IQS-8100 Series Transport Blazer

Typical Test Cases

4c. For STS-48c,

STS-192c,
AU-4-16c, and
AU-4-64c, select
the timeslot from
the Grid tab.

4d. For STS-768c, and

AU-4-256c
mapping, the
timeslot is
automatically
selected in the
Grid tab.

4e. Press Next or Finish.

5. Pattern Configuration:

5a. Set the pattern parameters. Refer to

Pattern TX on page 405 and Pattern
RX
on page 409 for more
information.

6. Press Finish to complete the test setup.

The Grid tab closes and automatically
switch to the Alarm summary tab.

7. For additional configuration parameters

and results, refer to Summary Tabs on page 125, Port Tabs on page 137,
SONET Tabs on page 211, SDH Tabs on page 303, BERT Tabs on
page 405,
Advanced Tabs on page 411, and Common Tabs on
page 491. F
or Unframed mode, only Summary, Port, and BERT tabs
are available.

8. Press the Start button to start the test. Refer to Global Test Status and

Controls on page 35 for more information on test management.