Ethernet Tabs
400
IQS-8100 Series Transport Blazer
Error/Alarm RX
Error Analysis
³
FCS: The number of received frames with an invalid FCS.
³
Jabber/Giant: The number of received frames larger than 1518 (no
VLAN tag), or 1522 (1 VLAN tag) bytes with an invalid FCS.
³
Oversize: The number of received frames larger than 1518 (no VLAN
tag), or 1522 (1 VLAN tag) bytes with a valid FCS. Oversize error
analysis is only available when Oversize Monitoring is enabled (see
page 399).
³
Runt: The number of received frames that are smaller than 64 bytes
with an invalid FCS.
³
Undersize: The number of received frames smaller than 64 bytes with
a valid FCS.
For Gb Ethernet only:
³
Symbol: A Symbol Error is declared when an invalid code-group in the
code is detected.
³
Idle: An Idle Error is declared when an error is detected between the
end of a frame and the beginning of the next frame.
³
False Carrier: A False Carrier is declared when data is being received
with no valid start of frame.
For Ethernet and 10G Ethernet only:
³
Block: The number of frames received with an errored block
condition. Not available with 10G Ethernet in GFP.
Total Error Count: Indicates the total number of errors including all the
above errors at the exception of Oversize when the Oversize Monitoring
check box is not selected.